33
GA--A 202
6.6 System messages
.
IMPORTANT!
If you cannot eliminate the fault or error with the help of this table, please
contact t he service department or return the device for repair.
' Do not attempt to do any repairs yourself!
L Depending on the status or error state the POWER CONTROL displays
the following messages on the RIWO--NET menu monitor:
6.6.1 Status messages
Type of message Message text Possible cause Corrective action
Status message 1 Motor of “PowerStick” blocked! Motor overloaded or tool blocked
'Interrupt application briefly /
check tool
Status message 2 No “PowerStick“ connected! No “PowerStick“ connected
'Connect PowerStick
Status message 3 No “Footswitch” connected! No “Footswitch” connected
'Connect footswitch
Status message 4 Lower limit of range is reached.
Lower limit of range of adjustable
parameters is reached
Status message 5 Upper limit of range is reached.
Upper limit of range of adjustable
parameters is reached
Status message 6 Motor of „PowerStick“ blocked!
Status message 7 Control level B activated
Status message 8 Control level A activated
“+ voice output” = The system messages marked with this symbol are accompanied by an accustic message
(only if the “Voice output” option is available).
6.6.2 Warnings
Type of message Message text Possible cause Corrective action
Warning 1 Caution! Footswitch actuated or defective
Footswitch is actuated when
connecting the cable or defective
'Let go of footswitch
'Contact service department
Warning 2 Caution! “shaver” device error Device malfunction/defect
'Contact service department
6.6.3 Errors
Errors Possible cause Corrective action
No log--in of device in the RIWO NET SYSTEM
Rotary coding switch “ID” not in
“0” position
'Set rotary coding switch “ID” to “0”