EasyManuals Logo

Roche Cedex Bio System User Manual

Roche Cedex Bio System
496 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #144 background imageLoading...
Page #144 background image
Roche Diagnostics
B-14 Operatorโ€™s Manual ยท Version 3.1
5 Daily operation Cedex Bio System
Preparing the system
5
Prepare the reagents. Deal with at least all red and yellow buttons.
1. Press a reagent button to check the details.
Replace the empty reagents.
2. Press .
3. Open the main cover
4. Remove the bottle.
5. Press to confirm.
6. Remove the remaining bottles of the set.
7. Press .
8. Place the new bottle.
9. Press to confirm.
10. Load the remaining bottles of the set.
11. Press to proceed to the next stage in the
Prepare wizard.
6
Prepare the cuvettes. Replace all red cuvette segments.
1. Press the cuvette button.
2. Open the main cover.
3. Replace the cuvette segment.
4. Press to confirm the replacement.
5. Replace the remaining segments that need
replacing.
6. Press to proceed to the next stage in the
Prepare wizard.
7
Perform reagent mixing. 1. Select the test(s).
2. Press .
8
Perform calibrations. Perform the calibrations.
1. Check the test selection.
2. Press .
3. Place the calibrators.
4. On the placement screen, press .
5. Press .
6. Validate the results.
Step User action
Table B-2 Steps for preparing the system

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals