B-14 Operator’s Manual · Version 3.1
5 Daily operation Cedex Bio System
Preparing the system
5
Prepare the reagents. Deal with at least all red and yellow buttons.
1. Press a reagent button to check the details.
Replace the empty reagents.
2. Press .
3. Open the main cover
4. Remove the bottle.
5. Press to confirm.
6. Remove the remaining bottles of the set.
7. Press .
8. Place the new bottle.
9. Press to confirm.
10. Load the remaining bottles of the set.
11. Press to proceed to the next stage in the
Prepare wizard.
6
Prepare the cuvettes. Replace all red cuvette segments.
1. Press the cuvette button.
2. Open the main cover.
3. Replace the cuvette segment.
4. Press to confirm the replacement.
5. Replace the remaining segments that need
replacing.
6. Press to proceed to the next stage in the
Prepare wizard.
7
Perform reagent mixing. 1. Select the test(s).
2. Press .
8
Perform calibrations. Perform the calibrations.
1. Check the test selection.
2. Press .
3. Place the calibrators.
4. On the placement screen, press .
5. Press .
6. Validate the results.
Step User action
Table B-2 Steps for preparing the system