Cedex Bio System 6 Special operations
Lot handling
A screen is displayed that lists the lot IDs for the used reagent set, calibrators, and
controls. (The following figure shows an example of sample-result lot-data.)
2
Scroll to display the information, if required.
3
Press to close the screen.
Adding a new lot
Lot data are usually entered manually from the package insert. With the exception of
reagent and diluent lots: they can be defined by scanning their barcodes only.
Preparation tasks Make sure that the controls and calibrators are assigned to the application.
e
See Preparing applications on page B-140.
a To define a control lot
1
Choose Workplace > Lot Data.
2
Press Controls.
A screen is displayed that lists all installed control lots.
3
Press .
A screen is displayed, asking you to scan the barcode transfer sheet or to type the
data manually.
Defining lots manually requires utmost diligence
During lot definition you need to type data that are directly relevant to result generation.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals