Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.1 B-9
Cedex Bio System 5 Daily operation
Introduction
Working with a host system
Communication with the instrument is defined during installation.
e
For an overview on working with a host system, see Host connectivity on page A-32.
9
Finishing the shift 1. Check for unfinished orders. Workplace > Orders
Choose > Not Finished
2. Check for non-validated results. Workplace > Result Review
Choose > Not Accepted
3. Check for non-transmitted results.
(If working with a host system only.)
Workplace > Result Review
Choose > Not Sent to Host
4. Start the End Shift wizard. Overview > End Shift
5. Perform the daily backup. Utilities > Export > Database
6. Export the full results. Utilities > Export > Results
7. Clean up the database. Workplace > Orders >
Workplace > Result Review>
Workplace > QC Status >
Workplace > QC History >
Workplace > Calibrations >
8. Perform the maintenance actions
that are due.
Utilities > Maintenance
9. Replace cuvettes. Overview >
10. Check the external fluid containers. Overview > >
11. Remove the reagent disk (if last
shift).
Overview > >
12. Log off the system. Overview > button with your user name
13. Shut down the system and switch off
the instrument (if last shift).
n/a n/a
Task Steps Navigation
With wizard As individual steps
Table B-1 Overview of the daily operation tasks

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals