5 Daily operation Cedex Bio System
Finishing the shift
Performing the daily backup
The Cedex Bio instrument can store orders and result data for one working day. It is
therefore necessary to export to an external medium all data that you need to keep.
During a database export, the full content of the database is copied to the USB stick.
The database data can be restored to the instrument if required (Utilities > Import >
Database).
e
For information on restoring the database, see Importing a database on page B-113.
a To perform the daily backup
1
Press .
A screen is displayed, asking you to insert the USB stick.
2
Insert the USB stick.
A screen is displayed showing the directory structure of the USB stick.
3
Select the directory where you want the backup file to be copied.
4
Press to confirm the selection.
The data are copied to the stick.
Database files names have the following name format:
dba_yyyymmddhhmmss.tgz.
5
Press to proceed to the next stage in the End Shift wizard.
A screen is displayed for deleting sample results.
When performing backup without using the End Shift wizard:
Choose Utilities > Export > Database.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals