EasyManua.ls Logo

Roche Cedex Bio System - Prime ISE Reference and Calibrator

Roche Cedex Bio System
496 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics
E-48 Operator’s Manual · Version 3.1
14 ISE maintenance Cedex Bio System
ISE maintenance actions
Prime ISE reference and calibrator
f Utilities > Maintenance > Prime ISE Reference and Calib.
This maintenance action ensures that no traces of a previous ISE Calibrator
indirect/urine and ISE Reference Solution are left in the tubing and so prevents drift
in the results.
Operator time Approximately 1 minute.
Prerequisites The system must be in Standby status.
Tools and materials required None
Hazards and precautions None
a To prime the ISE calibrator and ISE Reference Solution
1
Select the maintenance action Prime ISE Reference and Calib.
2
Press .
The maintenance definition screen is displayed.
3
Press to start the action.
The instrument pumps first ISE Reference Solution and then ISE Calibrator
indirect/urine through the ISE tubing and past the electrodes.
A message will inform you when the maintenance action is complete.
4
Press .

Table of Contents

Related product manuals