C-10 Operatorโ€™s Manual ยท Version 3.1
8 General maintenance Cedex Bio System
Maintenance actions
Deproteinize the probe
f Utilities > Maintenance > Deproteinize Probe
To maintain the efficiency of the instrument, you must clean and deproteinize the
probe regularly to prevent the build-up of contaminants.
Operator time Approximately 5 minutes.
Prerequisites The system must be in Standby status.
Tools and materials required m ISE Deproteinizer
m Activator
a To deproteinize the probe
1
Prepare one tube with ISE Deproteinizer and one with Activator.
2
Select the maintenance action Deproteinize Probe.
3
Press .
The maintenance definition screen is displayed.
4
Press .
A placement list is displayed.
5
Place the ISE Deproteinizer and Activator on the positions indicated on the
screen.
6
Press to confirm the placing.
The system first deproteinizes and then activates the probe.
A message will inform you when the maintenance action is complete.
7
Remove the tubes.
8
Press .
Make sure that you have read and understood section Safety information on page C-8.
The following warning messages in particular are relevant:
o Injury through working solutions on page C-8.
o Incorrect results due to build-up of contaminants on page C-8.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals