6 Special operations Cedex Bio System
Removing condensation water from the reagent cooler
Removing condensation water from the reagent cooler
a To remove water from the reagent cooler
1
Make sure the system is in Standby status.
2
Choose Overview > > .
3
Open the main cover.
4
Remove the reagent disk.
Be sure not to tilt it and place it on an even surface.
5
Wipe the inside of the reagent cooler with a cloth or paper to remove the water.
6
Insert the reagent disk.
7
Close the cover.