13 ISE operation Cedex Bio System
Daily operation
Preparing the system
Short guide
The following table provides an overview on the steps that make up the preparation
process.
e
For non-ISE-specific tasks, see Preparing the system on page B-12.
Step User action
1
Start the Prepare wizard. 1. On the Overview tab, press Prepare.
2
Check the status of the external
containers.
1. Refill the water container, if required, and con-
firm by pressing the water button. (If you refill
the water, empty the waste as well.)
2. Empty the waste container, if required, and
confirm by pressing the waste button.
3. Replace the cleaner bottle, if required, and con-
firm by pressing the cleaner button.
4. Press to proceed to the next stage in the
Prepare wizard.
3
Check the onboard ISE fluids. 1. Replace the ISE fluid bottles, if required.
2. Replace the electrodes, if required.
3. Press to proceed to the next stage in the
Prepare wizard.
4
Perform the maintenance
actions that are due.
1. Perform the Daily Prepare Actions mainte-
nance action.
2. Perform the other ISE maintenance actions
that are due.
Perform at least all red maintenance actions.
Table E-4 Steps for preparing the system

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals