5 Daily operation Cedex Bio System
Finishing the shift
Dealing with the external fluid bottles
Emptying the waste container
To prevent unpleasant smells and contamination of the environment, you need to
empty the waste at the end of your shift.
a To empty the waste container
1
Remove the tubing adapter from the yellow waste container and insert it in the
spare container.
2
Remove the bottle from the rack and place it on a firm and even surface.
3
Place the spare container on the rack.
4
Follow the instructions given in To check the waste container on page B-17.
Refilling the external water bottle
a To refill the external water container
1
Remove the tubing adapter from the white water container and place it on a clean
surface.
2
Follow the instructions given in To check the water container on page B-16.
Make sure that you have read and understood section Safety information on page B-5. The
following warning messages in particular are relevant:
o Injury through reagents and other working solutions on page B-5.
o Infection by biohazardous materials on page B-5.
o Infection by waste solution on page B-5.
Roche recommend to refill the water container whenever you empty the waste container.
When emptying the waste without using the End Shift wizard:
Choose Overview > > .
Danger of poor measurement quality due to inadequate water quality
Inadequate water quality may lead to incorrect results. Always use purified water of the
quality specified in section Technical specifications.
Roche recommend to empty the waste container whenever you refill the water container.
When refilling the water without using the End Shift wizard:
Choose Overview > > .