Cedex Bio System 14 ISE maintenance
ISE maintenance actions
Activate electrodes
f Utilities > Maintenance > Activate Electrodes
The electrode surfaces must be activated with fresh Activator so that the correct
potentials are measured. Electrodes may be damaged if they are deprived of regular
contact with Activator.
Operator time Approximately 4 minutes.
Prerequisites The system must be in Standby status.
Tools and materials required m Activator
a To activate the electrodes
1
Select the maintenance action Activate Electrodes.
2
Press .
The maintenance definition screen is displayed.
3
Press to start the action.
A placement list is displayed.
4
Place the Activator on the sample area position indicated on the screen.
5
Press to confirm the placement and to start the action.
A message will inform you when the maintenance action is complete.
6
Remove the Activator from the sample area.
7
Press .