Roche Diagnostics
B-24 Operatorโs Manual ยท Version 3.1
5 Daily operation Cedex Bio System
Preparing the system
a To check the status of individual reagent sets
1
Choose Overview > .
2
Check the colors of the reagent set buttons.
Interpreting the colors
A Press to add a reagent set.
B Press to display the disk content in a table.
C The number indicates the number of free
positions on the disk.
Figure B-5
Color Meaning Possible actions
No tests can be performed with this
reagent set.
The number of available tests is 0. Replace the reagent set.
The set is incomplete. Add the missing reagent.
The test needs calibrating. Perform calibration.
There are fewer than 10% of tests left
for this set.
Load a new reagent set as soon as pos-
sible.
The expiration date has passed.
A QC is due or its result has not been
accepted.
Perform QC as soon as possible or
accept the result
A more recent version of the applica-
tion has been imported.
Install the new version of the applica-
tion to start working with it.
This development channel test is run
without extra wash cycles.
Activate the extra wash cycle.
Roche strongly recommends to always
use extra wash cycles with develop-
ment channel applications and also to
always load extra cleaner when tests
with extra wash cycles are used.
Ready for use. No action is currently required.
There is no application that uses this
reagent set.
Add the missing application.
A required diluent or cleaner is not
on board.
Load the missing fluid.