Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.1 B-79
Cedex Bio System 5 Daily operation
Finishing the shift
6
Perform maintenance actions. 1. Check which maintenance actions are due.
2. Perform the maintenance actions.
Perform at least all red maintenance actions.
3. Press to proceed to the next stage in the
End Shift wizard.
7
Check the cuvette status. Replace at least all red cuvette segments.
1. Press the cuvette button.
2. Open the main cover.
3. Replace the cuvette segments.
4. Close the main cover.
5. Press to proceed to the next stage in the
End Shift wizard.
8
Remove the reagent disk. 1. Choose Overview > .
2. Press
1. Open the main cover.
2. Lift the disk from the instrument.
3. Place the disk in its container.
4. Press .
5. Close the main cover.
6. Store the reagent disk in a cool place.
9
Empty the waste container. 1. Empty the waste container.
2. Press the Waste button to confirm.
3. Refill the water container.
4. Press the Water button to confirm.
10
Finish your shift. 1. Log off the system.
2. Shut down the system and switch off the main
instrument. (If there is no other shift.)
Step User action
Table B-9 Steps for finishing the shift

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals