Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.1 B-151
Cedex Bio System 7 Configuration
Applications
4
Define the lot values. Do one of the following:
Ifโ€ฆ Do thisโ€ฆ
The lot data are
available from a bar-
code
1. Scan the barcode.
The screen is displayed again that lists all installed calibrator lots.
The new lot is selected.
You want to type the
data manually
1. Press .
2. Type the name (up to 10 alphanumeric characters) and press .
3. Type the material code and press .
Refer to the package insert.
4. Type the lot ID (up to nine alphanumeric characters) and press .
5. Type the expiration date and press .
Use the date format as indicated on the screen.
6. Type the number of calibrators (cups) you need to place on the
instrument. Use the information given on the package insert.
(Exclude calibrators that use system water from this number. See
step 14.)
7. Press to confirm the definitions.
8. Press *Add new test*.
9. Press .
A screen is displayed that shows all tests that have this calibrator
defined for them and that are not yet assigned to this lot.
10. Press one of the test buttons.
11. Press .
12. Define the first calibration value (target value).
13. Define the next calibration value.
You can define up to six calibration values, and you must define
them in descending order.
14. Select the value for Last is Water.
Press On if you want to calibrate with system water as zero cali-
brator. (In this case, no cup needs to be placed on the system. (See
step 7).
Press Off if you want to use a special zero calibrator for the cali-
bration. (The special zero calibrator needs to be placed on the
sample area.)
15. Press to confirm the definitions.
Calibration values (target values) in method sheets may be defined in increasing order of
concentration. Always define the values in decreasing order on the Cedex Bio instrument.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals