Roche Diagnostics
B-162 Operatorโs Manual ยท Version 3.1
7 Configuration Cedex Bio System
Configuration
Sample Barcode On, Off You can work with or without barcodes on sample
tubes. Use Off only if you exclusively work without
barcodes.
If you use On, during order definition the option to
type the sample ID manually is skipped
Order Mode Easy You work with one test panel on the test selection
screen, containing up to 25 tests.
Typical situation for this mode: You work with one
or two reagent disks, manual order identification,
identical sample and order IDs, possibly but not nec-
essarily with host connection.
Full You can work with up to six test tabs on the test selec-
tion screen, each containing up to 20 tests.
Typical situation for this mode: You distribute the
reagents across several (up to eight) reagent disks and
you predominantly work with specific groups of tests.
You also work with a host connection with automatic
order identification and result upload.
Sample ID
Handling
Order ID =
Sample ID
The order ID is automatically defined when you iden-
tify the sample.
Independent IDs You define the sample ID and order ID indepen-
dently of each other.
Grouped Sample
ID
Order and sample ID are identical.
For a given sample, the same sample ID is used all the
time. A three digit number is appended and automat-
ically increased with each new order.
Examples: 789-001 for the first order, 789-002 for the
second order, and so on.
Sample ID
Separator
One
alphanumeric
character
This definition is used in combination with Grouped
Sample ID. It defines the character that separates the
sample ID from the automatically appended number.
Examples: 789-001, 789.001, 789_001.
Auto Order-ID On, Off If you use On, the system automatically increments
the order ID number by one whenever you define a
new order.
If you use On, choose Off for Order ID = Sample ID.
Auto Print
Results
On, Off If you use On, sample results are printed as soon as all
results of the order are available and accepted.
Test Tab 1... 6
Name
Up to four
alphanumeric
characters
Display name for the test tab 1... 6
Test tabs are used to group tests that are usually used
together.
You can assign a test to more than one tab.
Test Tab 1... 6
Visible
On, Off Use On to make available the tab on the test selection
screen.
Item Values Comment
Table B-20 Workflow definitions