Cedex Bio System 8 General maintenance
Maintenance actions
6
Dispose of any remaining water.
7
Pour 500 mL hypochlorite solution in the water container.
8
Clean the water container with a brush.
9
Dispose of the solution.
10
Clean the interior of the water container manually using tissues. Remove any
residue.
11
Rinse the water container with tap water at least four times.
12
Rinse the water container with purified water.
13
Wipe the suction pipe with a tissue.
14
Fill the water container with purified water.
15
Reinsert the tubing adapter into the water container. Press down firmly.
a To clean the waste container
1
Have the spare waste container ready.
The system periodically performs wash actions. Therefore, an external waste
container must be connected at all times.
2
Remove the tubing adapter from the yellow waste container and insert it in the
spare container.
3
Remove the bottle from the rack and place it on a firm and even surface.
4
Place the spare container on the rack.
5
Check that the tubing adapter is properly inserted.
6
Empty the removed waste container. Treat the fluid as biohazardous waste.
7
Pour 500 mL hypochlorite solution in the waste container.
8
Clean the container with a brush.
9
Dispose of the fluid. Treat it as biohazardous waste.
10
Clean the interior of the waste container manually using tissues. Remove any
residue.
11
Rinse the waste container with tap water.
12
Empty the container completely.