EasyManuals Logo

Roche Cedex Bio System User Manual

Roche Cedex Bio System
496 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #322 background imageLoading...
Page #322 background image
Roche Diagnostics
C-20 Operatorโ€™s Manual ยท Version 3.1
8 General maintenance Cedex Bio System
Maintenance actions
6
Replace the water inlet filter.
o Remove the retaining ring by pulling it by its tab away from the filter casing
(A, D, E).
o Remove the filter (B) and replace it with a new one.
o Place the retaining ring.
Press it firmly on the filter casing by placing your thumb across it an applying
equal pressure all around.
7
Reinsert the tubing adapter into the water container. Press down firmly.
8
Press .
The maintenance action Prime Fluid System is automatically performed.
A message will inform you when the maintenance action is complete.
9
Press .
A Removing the retaining ring
B Filter
C Filter casing
D Retaining ring
E Tab on retaining ring
Figure C-2
A
CB
E
D

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals