8 General maintenance Cedex Bio System
Maintenance actions
Preparing the instrument
13
Reinsert the sample tray.
14
Close the right service flap and the main cover.
15
Press .
The system automatically initializes the instrument.
A message will inform you when the maintenance action is complete.
16
Press .
17
Perform an air/water calibration.
You cannot process orders unless you perform this action.
e
See Perform an air/water calibration on page C-28.