Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.1 C-31
Cedex Bio System 8 General maintenance
Maintenance actions
6
Lift the tubing adapter from the white water container.
7
Place the tubing adapter on a clean surface.
Make sure the filter (A) is not obstructed and air can get to it.
8
Press to initialize the fluid sensor.
The system aspirates air through the tubing adapter and initializes the fluid
sensor.
When initialization is complete, you will be asked to insert the tubing adapter.
9
Insert the tubing adapter in the water container.
10
Press .
The fluid system is primed.
A message will inform you when the maintenance action is complete.
11
Press .
A