Roche Diagnostics
C-36 Operator’s Manual · Version 3.1
8 General maintenance Cedex Bio System
Maintenance actions
26
Install the sample tray.
Make sure to align the guides with the pins.
Press down firmly.
27
Install and close the right service flap.
28
Move the transfer head to its rightmost position.
29
Install and close the left service flap and main cover.
Finishing the maintenance
action
30
Switch on the instrument.
31
Wait for initialization to finish.
32
Log on to the system.
33
Insert the reagent disk.
e
See Preparing the reagent disk on page B-21