13 ISE operation Cedex Bio System
Daily operation
8
Performing controls
Performing
Default QC
1. Start the wizard. Overview > Order >
2. Select a control and place the tube.
Repeat until there are no controls left
on the screen.
n/a
3. Start the QC measurement.
4. Validate the QC results. Workplace > QC Status >
5. Remove the controls. n/a
Performing an individual
QC measurement
1. Start the wizard. Workplace > QC Status >
2. Select a test. n/a
3. Select a control and place the tube.
Repeat until there are no controls left
on the screen.
n/a
4. Start the QC measurement.
5. Validate the QC results. Workplace > QC Status >
6. Remove the control.
Performing all
due QC measurements
1. Start the wizard. Overview > Order >
>
2. Select a control and place the tube.
Repeat until there are no controls left
on the screen.
n/a
3. Start the QC measurement.
4. Validate the QC results. Workplace > QC Status >
5. Remove the controls.
9
Finishing the shift 1. Check for unfinished orders. Workplace > Orders
Choose > Not Finished
2. Check for non-validated results. Workplace > Result Review
Choose > Not Accepted
3. Check for non-transmitted results.
(If working with a host system only.)
Workplace > Result Review
Choose > Not Sent to Host
4. Start the End Shift wizard. Overview > End Shift
5. Perform the daily backup. Utilities > Export > Database
6. Export the full results. Utilities > Export > Results
7. Clean up the database. Workplace > Orders >
Workplace > Result Review>
Workplace > QC Status >
Workplace > QC History >
Workplace > Calibrations >
8. Perform the maintenance actions
that are due.
Utilities > Maintenance
9. Check the external fluid containers. Overview > >
10. Log off the system. Overview > button with your user name
11. Shut down the system and switch off
the instrument (if last shift).
n/a n/a
Task Steps Navigation
With wizard As individual steps
Table E-3 Overview of the daily operation tasks