Command Reference
R&S
®
CMW500
247User Manual 1173.9463.02 ─ 02
7.3.8.4 Address Settings
The following commands define the address information for the different remote-control
interfaces of the R&S CMW500.
SYSTem:COMMunicate:SOCKet<i>:MODE......................................................................247
SYSTem:COMMunicate:SOCKet<i>:PORT.......................................................................247
SYSTem:COMMunicate:SOCKet<i>:VRESource...............................................................248
SYSTem:COMMunicate:GPIB<i>[:SELF]:ADDR................................................................248
SYSTem:COMMunicate:GPIB<i>[:SELF]:ENABle..............................................................248
SYSTem:COMMunicate:GPIB<i>:VRESource...................................................................249
SYSTem:COMMunicate:USB:VRESource........................................................................249
SYSTem:COMMunicate:VXI<i>:GTR...............................................................................250
SYSTem:COMMunicate:VXI<i>:VRESource.....................................................................250
SYSTem:COMMunicate:SOCKet<i>:MODE <Mode>
Sets the protocol operation mode for direct socket communication.
Suffix:
<i>
.
1 to 2
Selects the remote channel.
Parameters:
<Mode> RAW | AGILent | IEEE1174
RAW: no support of control messages, e.g. polling or service
request
AGILent: emulation codes via control connection (control port)
IEEE1174: emulation codes via data connection (data port
*RST: a *RST has no effect on this parameter
Example:
SYSTem:COMMunicate:SOCKet:MODE RAW
Set the operation mode to raw TCP/IP.
Firmware/Software: V1.0.5.3
SYSTem:COMMunicate:SOCKet<i>:PORT <Port>
Sets the port number for direct socket communication.
Suffix:
<i>
.
1 to 2
Selects the remote channel.
Parameters:
<Port> Range: 1024 to 32767
*RST: a *RST has no effect on the value
Example:
SYSTem:COMMunicate:SOCKet:PORT 1025
Set the port number to 1025 (channel 1).
Firmware/Software: V1.0.5.3