Basic Instrument Functions
R&S
®
CMW500
86User Manual 1173.9463.02 ─ 02
The instrument setup
SCPI command:
SYSTem:PRESet etc.
Reset
Sets the instrument parameters to values suitable for remote operation. In particular, the
"Reset" state comprises the following settings:
All measurements are performed in Single-Shot mode ("Repetition: Single-Shot").
The R&S CMW500 uses short statistics cycles (for benchmarks).
In multi evaluation measurements, some time-consuming evaluations are skipped to
gain measurement speed.
The preferred modes and features are the same in all software versions. As far as
older features are concerned, the Reset state never changes after a software
upgrade.
Exceptions:
The following R&S CMW500 settings are not affected by "Reset":
Address information assigned to the instrument (e.g. the IP address)
The instrument setup
The contents of the status registers
SCPI command:
SYSTem:RESet etc.
Factory Default
Sets the instrument to its factory default state. This state comprises the following settings:
All Preset settings: The instrument is optimized for local/manual operation.
The address information assigned to the instrument (e.g. the IP address or USB
product ID/vendor ID/device ID).
The instrument setup: One sub-instrument.
Use the Startup Dialog to restore the factory default settings.
SCPI command:
n/a
5.2.2 Save/Recall Dialog
The "Save/Recall" dialog stores the current instrument setup and recalls (and re-acti-
vates) setups.
To open the dialog press SAVE.
Utility Dialogs