Getting Started
R&S
®
CMW500
52User Manual 1173.9463.02 ─ 02
3. Press the TASKS key and select the "GPRF Meas Power" measurement from the
task bar across the bottom of the display (see also Configuring the Task Bar).
The "General Purpose RF - Power" dialog is opened.
4. Press ON / OFF (or RESTART / STOP) to start the measurement.
5. Press "RF Settings > RF Routing..." and ensure that RF2 COM is selected as an input
connector.
6. In the "Settings" panel of the measurement dialog, adjust the "Frequency", the
"Expected Nominal Power", and the filter "Bandwidth" to the properties of your input
signal.
7. Observe the measurement result on the screen.
In the present example, the upper tone (at 1200.001 MHz) of the generated dual-tone
signal is observed in a 1 kHz bandwidth. Due to the cable loss, the measured power is
smaller than the 36.19 dBm shown in figure 3-1.
Fig. 3-3: RF Power Results
3.2.3 Performing Signaling Measurements
The purpose of a signaling application is to establish a network connection to a mobile
station in order to perform various tests.
Sample Session