Command Reference
R&S
®
CMW500
257User Manual 1173.9463.02 ─ 02
Example:
SYSTem:ERRor:CODE?
Query the oldest entry in the error queue. "0" is returned if the error
queue is empty.
Usage: Query only
SCPI confirmed
Firmware/Software: V1.0.0.4
SYSTem:ERRor:COUNt?
Queries the number of entries in the error queue.
Example:
SYSTem:ERRor:COUNt?
If the queue is empty, 0 is returned.
Usage: Query only
SCPI confirmed
Firmware/Software: V1.0.0.4
SYSTem:ERRor[:NEXT]?
Queries and at the same time deletes the oldest entry in the error queue. Operation is
identical to that of STATus:QUEue[:NEXT].
The entry consists of an error number and a short description of the error. Positive error
numbers are instrument-dependent. Negative error numbers are reserved by the SCPI
standard.
Example:
SYSTem:ERRor?
Query the oldest entry in the error queue. 0,"No error" is returned
if the error queue is empty.
Usage: Query only
SCPI confirmed
Firmware/Software: V1.0.0.4
SYSTem:HELP:STATus:BITS?
Returns a list of paths for the bits of the STATus:OPERation registers at the lowest level
of the hierarchy.
Each path is represented by a string containing all registers from highest to lowest level
separated by colons.
Example: "STATus:OPERation:TASK:A:GPRF:MEASurement:POWer:OFF"
Usage:
Query only
Firmware/Software: V2.0.10
Instrument-Control Commands