Command Reference
R&S
®
CMW500
265User Manual 1173.9463.02 ─ 02
Parameters:
<Direction> IN | OUT
IN: Input connector
OUT: Output connector
*RST: IN
Example:
TRIGger:BASE:EXTB:DIRection OUT
Use TRIG B as an output connector.
Firmware/Software: V1.0.4.11
TRIGger:
BASE:EXTA:SOURce '<Source>'
TRIGger:BASE:EXTB:SOURce '<Source>'
Selects the output trigger signals to be routed to the TRIG A and TRIG B connectors. The
signals listed below are always available. Depending on the installed options additional
values may be available. A complete list of all supported values can be retrieved using
TRIGger:...:CATalog:SOURce?.
Parameters:
'<Source>' 'No Connection': No output trigger signal available
'Base1: External TRIG A': Input trigger signal from the TRIG A
connector
'Base1: External TRIG B': Input trigger signal from the TRIG B
connector
*RST: 'No Connection'
Example:
TRIGger:BASE:EXTB:SOURce "Base1: External TRIG
A"
Select the input signal from TRIG A as an output trigger signal for
connector TRIG B.
Firmware/Software: V1.0.4.11
Instrument-Control Commands