If necessary, the command activates the delta marker and positions a reference
marker to the peak power.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<Position> Numeric value that defines the marker position on the x-axis.
Range: The value range and unit depend on the measure-
ment and scale of the x-axis.
Example:
CALC:DELT:X?
Outputs the absolute x-value of delta marker 1.
CALCulate<n>:DELTamarker<m>:Y
Suffix:
<n>
.
1..n
<m> 1..n
Return values:
<Result> Result at the position of the delta marker.
The unit is variable and depends on the one you have currently
set.
Default unit: DBM
CALCulate<n>:MARKer<m>:X <Position>
This command moves a marker to a specific coordinate on the x-axis.
If necessary, the command activates the marker.
If the marker has been used as a delta marker, the command turns it into a normal
marker.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<Position> Numeric value that defines the marker position on the x-axis.
The unit depends on the result display.
Range: The range depends on the current x-axis range.
Default unit: Hz
Example:
CALC:MARK2:X 1.7MHz
Positions marker 2 to frequency 1.7 MHz.
Manual operation: See " Marker Table " on page 17
See " Marker Peak List " on page 18