Remote Commands for the I/Q Analyzer
R&S
®
FSWP
103User Manual 1177.5856.02 ─ 07
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<SortMode> X
Sorts the peaks according to increasing position on the x-axis.
Y
Sorts the peaks according to decreasing position on the y-axis.
*RST: X
Example:
CALC:MARK:FUNC:FPE:SORT Y
Sets the sort mode to decreasing y values
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:X?
This command queries the position of the peaks on the x-axis.
The order depends on the sort order that has been set with CALCulate<n>:
MARKer<m>:FUNCtion:FPEaks:SORT.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Return values:
<PeakPosition> Position of the peaks on the x-axis. The unit depends on the
measurement.
Usage: Query only
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:Y?
This command queries the position of the peaks on the y-axis.
The order depends on the sort order that has been set with CALCulate<n>:
MARKer<m>:FUNCtion:FPEaks:SORT.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Return values:
<PeakPosition> Position of the peaks on the y-axis. The unit depends on the
measurement.
Usage: Query only
CALCulate<n>:DELTamarker<m>:X <Position>
This command moves a delta marker to a particular coordinate on the x-axis.
Retrieving Results