Remote Commands for the I/Q Analyzer
R&S
®
FSWP
107User Manual 1177.5856.02 ─ 07
Parameters:
<FileName> String containing the path and name of the source file. The file
extension is *.iq.tar.
<AnalysisBW> Optionally: The analysis bandwidth to be used by the measure-
ment. The bandwidth must be smaller than or equal to the band-
width of the data that was stored in the file.
Default unit: HZ
Example:
INP:FILE:PATH 'C:\R_S\Instr\user\data.iq.tar'
Uses I/Q data from the specified file as input.
Manual operation: See " Select I/Q data file " on page 45
INPut<ip>:FILTer:HPASs[:STATe] <State>
Activates an additional internal high-pass filter for RF input signals from 1 GHz to
3 GHz. This filter is used to remove the harmonics of the R&S FSWP in order to mea-
sure the harmonics for a DUT, for example.
This function requires an additional high-pass filter hardware option.
(Note: for RF input signals outside the specified range, the high-pass filter has no
effect. For signals with a frequency of approximately 4 GHz upwards, the harmonics
are suppressed sufficiently by the YIG-preselector, if available.)
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST: 0
Example:
INP:FILT:HPAS ON
Turns on the filter.
Manual operation: See " High Pass Filter 1 to 3 GHz " on page 43
INPut<ip>:FILTer:YIG[:STATe] <State>
Suffix:
<ip>
.
1 | 2
irrelevant
Example:
INP:FILT:YIG OFF
Deactivates the YIG-preselector.
Manual operation: See " YIG-Preselector " on page 43
Configuring I/Q Analyzer Measurements