Remote Commands for the I/Q Analyzer
R&S
®
FSWP
106User Manual 1177.5856.02 ─ 07
INPut<ip>:IMPedance.....................................................................................................108
INPut<ip>:IMPedance:PTYPe......................................................................................... 108
INPut<ip>:SELect...........................................................................................................108
INPut<ip>:UPORt:STATe.................................................................................................109
INPut<ip>:UPORt[:VALue]...............................................................................................109
TRACe:IQ:FILE:REPetition:COUNt.................................................................................. 110
INPut<ip>:ATTenuation:PROTection:RESet
This command resets the attenuator and reconnects the RF input with the input mixer
for the R&S FSWP after an overload condition occurred and the protection mechanism
intervened. The error status bit (bit 3 in the STAT:QUES:POW status register) and the
INPUT OVLD message in the status bar are cleared.
The command works only if the overload condition has been eliminated first.
Suffix:
<ip>
.
1 | 2
irrelevant
Example:
INP:ATT:PROT:RES
INPut<ip>:COUPling <CouplingType>
This command selects the coupling type of the RF input.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<CouplingType> AC | DC
AC
AC coupling
DC
DC coupling
*RST: AC
Example:
INP:COUP DC
Manual operation: See " Input Coupling " on page 42
INPut<ip>:FILE:PATH <FileName>[, <AnalysisBW>]
This command selects the I/Q data file to be used as input for further measurements.
The I/Q data must have a specific format as described in Chapter A.3, "I/Q Data File
Format (iq-tar)", on page 154.
Suffix:
<ip>
.
1 | 2
irrelevant
Configuring I/Q Analyzer Measurements