Suffix:
<ip>
.
1..n
Parameters:
<Source> RF
Radio Frequency ("RF INPUT" connector)
FIQ
I/Q data file
(selected by INPut<ip>:FILE:PATH on page 106)
*RST: RF
Manual operation: See " Radio Frequency State " on page 42
See "I/Q Input File State" on page 44
INPut<ip>:UPORt:STATe
<State>
This command toggles the control lines of the user ports for the AUX PORT connector.
This 9-pole SUB-D male connector is located on the rear panel of the R&S FSWP.
See the R&S FSWP Getting Started manual for details.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State> ON | 1
User port is switched to INPut
OFF | 0
User port is switched to OUTPut
*RST: 1
INPut<ip>:UPORt[:VALue]
This command queries the control lines of the user ports.
For details see OUTPut<up>:UPORt[:VALue] on page 110.
Suffix:
<ip>
.
1 | 2
irrelevant
Return values:
<Level> bit values in hexadecimal format
TTL type voltage levels (max. 5V)
Range: #B00000000 to #B00111111
Example:
INP:UPOR?
//Result: #B00100100
Pins 5 and 7 are active.