Configuration
R&S
®
FSWP
68User Manual 1177.5856.02 ─ 07
Remote command:
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer on page 138
Lower Level Hysteresis
When the reference level is adjusted automatically using the Auto Level function, the
internal attenuators and the preamplifier are also adjusted. To avoid frequent adapta-
tion due to small changes in the input signal, you can define a hysteresis. This setting
defines a lower threshold the signal must fall below (compared to the last measure-
ment) before the reference level is adapted automatically.
Remote command:
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer on page 138
Adjusting Settings Automatically