R&S ZVL GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
220
One-port calibrations can be performed automatically or manually.
• Calibration Unit 75 Q starts an automatic full one-port calibration. This function is enabled while
the Calibration Unit is connected.
The following menu commands calls up the start dialog of the calibration wizard to start a manual
calibration:
• Normalize (Short) initiates a normalization using a short standard.
• Full initiates a full one-port calibration.
Remote
control:
[SENSe<Ch>:]CORRection:COLLect:AUTO ' ', <port_no> (for matching
analyzer and cal unit ports)
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs (for arbitrary pairs of ports)
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
Two-Port P1 P2
Opens a submenu to select a two-port calibration at test ports PORT 1 and PORT 2:
One-port calibrations can be performed automatically or manually.
• Calibration Unit 75 Q starts an automatic tow-port TOSM calibration. This function is enabled
while the Calibration Unit is connected.
The following menu commands calls up the start dialog of the calibration wizard to start a manual
calibration:
• Normalization Forward... initiates a normalization in forward direction (using PORT 1 as a source
port and a through standard). This calibration type is preferable (compared to a normalization in
both directions), if only forward transmission parameters (S
21
) are measured.
• Normalization Backward... initiates a normalization in backward (reverse) direction (using PORT 2
as a source port and a through standard). This calibration type is preferable (compared to a
normalization in both directions), if only reverse transmission parameters (S
12
) are measured.
• Normalization Both Directions... initiates a normalization in forward and backward direction (using