R&S ZVL GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
221
a t
hrough standard). This calibration type is preferable (compared to a unidirectional
nor
malization), if both transmission parameters (S
21
and S
12
) are measured.
One Path Two-Port initiates a full one-port calibration.
TOSM initiates a Through-Open-Short-Match (12-term) calibration
Remote
control:
[SENSe<Ch>:]CORRection:COLLect:AUTO ' ', <port_no> (for matching
analyzer and cal unit ports)
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs (for arbitrary pairs of ports)
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
Guided Calibration
The analyzer provides a calibration wizard for each calibration type. The guided calibration consists of the
following steps:
Select Physical Port Connectors and calibration kits at all calibrated ports.
Measure Standards: Acquire measurement data for all standards required for the selected calibration
type.
Calculate the system error correction data (error terms) from the measurement data of the standards
and apply the result to the active channel.
A successful calibration will supersede the previous calibration, discarding all previous system error
correction data. To keep older correction data you can transfer them into a Cal Pool using the Calibration
Manager.
The system error correction data determined in a calibration procedure are stored on the analyzer. You
can read these correction data using the remote control command [SENSe<Ch>:]CORRection:CDATa.
You can also replace the correction data of the analyzer by your own correction data sets.
Select Connectors
The first dialog of the calibration wizard displays a table to select the connectors and calibration kits for all
calibrated physical ports.