R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
379
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER:STARt <numeric_value>
Defines the start value of the search range selected with
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <numeric_value>.
<Chn>
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10.
<numeric_value>
Beginning of the search range.
Range [def. unit] Maximum allowed sweep range, depending on the instrument model and on
the sweep type. [Hz, dBm or s, depending on the sweep type]
*RST value 0 Hz
SCPI, Command
Types
Device-specific, command or query.
Example:
See CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <numeric_value>
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER:STOP <numeric_value>
Defines the stop value of the search range selected with
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <numeric_value>.
<Chn>
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10.
<numeric_value>
End of the search range.
Range [def. unit] Maximum allowed sweep range, depending on the instrument model and on
the sweep type. [Hz, dBm or s, depending on the sweep type]
*RST value 0 Hz
SCPI, Command
Types
Device-specific, command or query.
Example:
See CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <numeric_value>
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute MAXimum | MINimum | RPEak
| LPEak | NPEak | TARGet | LTARget | RTARget | BFILter
Selects a search mode for marker no. <Mk> and initiates the search. The marker must be created before
using CALCulate<Chn>:MARKer<Mk>[:STATe] ON (exception: bandfilter search).
<Chn>
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10. For a bandfilter search (BFILter) this
numeric suffix is ignored and may be set to any value because the bandfilter
search functions always use markers Mkr1 to Mkr4.
Parameters
See list of parameters below.
*RST value
SCPI,
Command
Types
Device-specific, no query.
Example: