EasyManua.ls Logo

Rohde & Schwarz ZVL series - Page 425

Rohde & Schwarz ZVL series
624 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
410
Use CALCulate<Chn>:TRANsform:TIME:LPASs if you wish to use one of the other algorithms for
calculating the grid.
<Chn>
Channel number used to identify the active trace
Parameters
None
*RST value
SCPI, Command Types
Device-specific, no query (the command causes an event)
Example:
See CALCulate<Chn>:TRANsform:TIME[:TYPE]
CALCulate<Chn>:TRANsform:TIME:SPAN <numeric_value>
Defines the time span of the diagram in time domain.
<Chn>
Channel number used to identify the active trace
<numeric_value>
Time span of the diagram in time domain.
Range [def. unit] 2E-012 s to 200 s. The increment (parameters UP or DOWN) is 0.1 ns. [s]
*RST value 5E-009 s
SCPI, Command Types
Confirmed, command or query
Example:
See CALCulate<Chn>:TRANsform:TIME:CENTer
Note: If the x-axis is scaled in distance units (CALCulate<Chn>:TRANsform:TIME:XAXis
DISTance), then the span is entered in m; the range and default value changes accordingly.
CALCulate<Chn>:TRANsform:TIME:STARt <numeric_value>
Defines the start time of the diagram in time domain.
<Chn>
Channel number used to identify the active trace
<numeric_value>
Start time of the diagram.
Range [def. unit] –100 s to +99.999999999998 s. The increment (parameters UP or DOWN) is
0.1 ns. [s]
*RST value –1E-009 s
SCPI, Command
Types
Confirmed, command or query
Example:
*RST; CALC:TRAN:TIME:STAT ON
Reset the instrument, activating a frequency sweep, and enable the time
domain transformation for the default trace.
CALC:TRAN:TIME:STAR 0; STOP 10 ns
Set the start time to 0 ns and the stop time to 10 ns.
Note: If the start frequency entered is greater than the current stop frequency
(CALCulate<Chn>:TRANsform:TIME:STOP), the stop frequency is set to the start frequency
plus the minimum frequency span (CALCulate<Chn>:TRANsform:TIME:SPAN).