R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
411
If the x-axis is scaled in distance units (CALC
ulate<Chn>:TRANsform:TIME:XAXis
DISTance), then the start value is entered in m; the range and default value changes accordingly.
CALCulate<Chn>:TRANsform:TIME:STATe <Boolean>
Determines whether the time domain transformation for trace no. <Chn> is enabled.
<Chn>
Channel number used to identify the active trace
<Boolean>
ON – Time domain representation active
OFF – Frequency domain representation active
*RST value OFF [–]
SCPI, Command
Types
Confirmed, command or query
Example:
*RST; CALC:TRAN:TIME:STAT?
Reset the instrument, activating a frequency sweep, and query whether the
default trace is displayed in the time domain. The response is 0.
CALCulate<Chn>:TRANsform:TIME:STIMulus STEP | IMPulse
Selects the type of stimulus to be simulated in the low pass transformation process.
<Chn>
Channel number used to identify the active trace
Parameters
IMPulse – Impulse response, in bandpass or lowpass mode
STEP – Step response, only in lowpass mode (a bandpass mode setting
CALCulate<Chn>:TRANsform:TIME[:TYPE] BPASs is automatically changed
to lowpass)
*RST value IMP
SCPI,
Command
Types
Confirmed, command or query
Example:
See CALCulate<Chn>:TRANsform:TIME[:TYPE]
CALCulate<Chn>:TRANsform:TIME:STOP <numeric_value>
Defines the stop time of the diagram in time domain.
<Chn>
Channel number used to identify the active trace
<numeric_value>
Stop time of the diagram.
Range [def. unit] –99.999999999998 s to +100 s. The increment (parameters UP or DOWN)
is 0.1 ns. [s]
*RST value +4E-009 s
SCPI, Command
Types
Confirmed, command or query
Example:
See CALCulate<Chn>:TRANsform:TIME:STARt.