R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
476
Example:
See MMEMory:LOAD:LIMit.
MMEMory:STORe:MARKer '<file_name>'
Saves the values of all markers to a ASCII file.
'<file_name>'
String parameter to specify the name and directory of the created ASCII file. The
default extension (manual control) for marker files is *.txt, although other
extensions are allowed. If no path is specified the analyzer uses the current
directory, to be queried with MMEMory:CDIRectory?. An example for the file
contents is given below.
*RST value
SCPI,
Command
Types
Device-specific, no query.
Example:
*RST
Reset the analyzer, creating the default trace no. 1 in
channel no. 1.
CALC:MARK ON; MARK:X 1GHz
Create marker no. 1 and place it to 1 GHz.
CALC:MARK2 ON; MARK2:X 2GHz
Create a second marker and place it to 2 GHz.
MMEM:STOR:MARK 'Marker.txt'
Store the marker values to an ASCII file. The file contains
both marker values:
MMEMory:STORe:NWANalyzer <numeric_value>,'<file_name>'
Stores the configuration data of the current setup to a specified setup file.
MMEMory:STORe:NWANalyzer renames the current setup, appending a .nwa extension. See
example for MMEMory:LOAD:NWANalyzer.
<numeric_value>
1 (the <numeric_value> is used for compatibility with the SCPI standard but is
ignored).
'<file_name>'
String parameter to specify the name and directory of the created setup file.
The default extension (manual control) for setup files is *.nwa, although other
extensions are allowed. If no path is specified the analyzer uses the current
directory, to be queried with MMEMory:CDIRectory?.
*RST value
SCPI, Command
Confirmed, no query.