<Ch>
Channel number of the offset-corrected channel
<port_no>
Port number of the analyzer
<delay>
Delay
Range [def. unit] -3.40282346638529E+038 s to +3.40282346638529E+038 s [s].
*RST value 0 s
SCPI, Command Types
Device-specific, command or query
Example:
See [SENSe<Ch>:]CORRection:EDELay<port_no>:ELENgth
[SENSe<Ch>:]CORRection:FACTory[:STATe] <Boolean>
Enables or disables the factory calibration for a particular channel.
<Ch>
Channel number of the calibrated channel
<Boolean>
Enable or disable factory calibration
*RST value ON
SCPI, Command
Types
Device-specific, command or query
Example:
*RST; CORR:FACT?
Reset the instrument and query whether the factory calibration for channel 1 is
enabled. The response is 1.
[SENSe<Ch>:]CORRection:LOSS<port_no> <ref_loss>
Defines the offset loss at the reference frequency
([SENSe<Ch>:]CORRection:LOSS<port_no>:FREQuency).
<Ch>
Channel number of the offset-corrected channel
<port_no>
Port number of the analyzer
<ref_loss>
Frequency-dependent part of the offset loss
Range [def. unit] -200 dB to +200 dB [dB]. The increment (UP/DOWN) is 0.001 dB.
*RST value 0 dB
SCPI, Command Types
Device-specific, command or query
Example:
See [SENSe<Ch>:]CORRection:EDELay<port_no>:ELENgth
[SENSe<Ch>:]CORRection:LOSS<port_no>:AUTO ONCE
Defines the offset parameters for the active test port such that the residual delay of the active trace
(defined as the negative derivative of the phase response) is minimized and the measured loss is
reproduced as far as possible across the entire sweep range.
<Ch>
Channel number of the offset-corrected channel.