R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
509
<port_no>
Port number of the analyzer. This numeric suffix is ignored; the active port is
determined by the active trace.
ONCE Applies the Auto Length and Loss function.
*RST value
SCPI, Command
Types
Device-specific, no query
Example:
*RST; CORR:LOSS:AUTO ONCE
Reset the instrument and apply the Auto Length and Loss function to the
default trace (Trc1 in channel 1).
[SENSe<Ch>:]CORRection:LOSS<port_no>:FREQuency <ref_frequency>
Defines the reference frequency for the frequency-dependent part of the offset loss
([SENSe<Ch>:]CORRection:LOSS<port_no>:OFFSet).
<Ch>
Channel number of the offset-corrected channel
<port_no>
Port number of the analyzer
<ref_frequency>
Reference frequency
Range [def. unit] Frequency range of the analyzer model [Hz]. The increment (UP/DOWN) is
1 MHz.
*RST value 1000000000 Hz (= 1 GHz)
SCPI, Command
Types
Device-specific, command or query
Example:
See [SENSe<Ch>:]CORRection:EDELay<port_no>:ELENgth
[SENSe<Ch>:]CORRection:LOSS<port_no>:OFFSet <DC_loss>
Defines the frequency-independent part (DC value) of the offset loss.
<Ch>
Channel number of the offset-corrected channel
<port_no>
Port number of the analyzer
<DC_loss>
Frequency-independent part of the offset loss
Range [def. unit] -200 dB to +200 dB [dB]. The increment (UP/DOWN) is 0.001 dB.
*RST value 0 dB
SCPI, Command Types
Confirmed, command or query
Example:
See [SENSe<Ch>:]CORRection:EDELay<port_no>:ELENgth
[SENSe<Ch>:]CORRection:OFFSet<port_no>[:STATe] <Boolean>
Resets the offset parameters for all test ports to zero or queries whether any of the offset parameters are
different from zero.
<Ch>
Channel number of the offset-corrected channel

Table of Contents

Related product manuals