Cond cm log Scale Scale
DISPLAY per per mult. base Factor Factor
UNITS TEMP. Siemens length 10 Byte Byte Byte
COMP. (dec) (hex) Rxd Expect
18 Kohm-m YES 0.001 0.01 0.00001 -5 251 FB 41 70
42 66
19
Kohm-m Raw
NO 0.001 0.01 0.00001 -5 251 FB 43 70
44 66
20 Ohm-m YES 1 0.01 0.01 -2 254 FE 45 70
46 69
21 Ohm-m Raw NO 1 0.01 0.01 -2 254 FE 47 70
48 69
22 Ohm-cm YES 11100004948
50 48
23
Ohm-cm Raw
NO 11100005148
52 48
24 %CONC YES 11100005348
54 48
25 spare n/a 0 00 55 48
56 48
26 ppb-TDS not supported 0 00 57 48
58 48
27 spare n/a 0 00 59 48
60 48
28 ppm-TDS YES 11100006148
62 48
29 spare n/a 11100006348
64 48
30 ppt-TDS YES 0.001 1 0.001 -3 253 FD 65 70
66 68
31 spare n/a 0 00 67 48
68 48
32 %SAT not supported 0 00 69 48
70 48
33 spare n/a 0 00 71 48
72 48
34pH YES 11100007348
74 48
35 mV-ORP NO 11100007548
76 48
36 NTU not supported 0 00 77 48
78 48
37 ppm-NTU not supported 0 00 79 48
80 48
38 DO not supported 0 00 81 48
82 48
39 DegC N/A 11100008348
84 48
A-18
MODEL 2700 APPENDIX A
SERIAL COMMUNICATIONS PROTOCOL
TABLE 4. Scale Bytes from Send Scale Factors command (continued)