Multi-Port Optical Power Meter MPM-210
5-5
1. ECHO
Description
Denite echo
Syntax
ECHO<wsp><value>
Parameters
0 : Disable
1 : Enable
Default
0
Example
ECHO 0
2. *IDN?
Description
Identication Query
Syntax
*IDN?
Response
ANTEC,MPM-210,zzzzzzzz,Ver. x.y
zzzzzzzz is serial number of main frame, and Ver. x.y is
firmware version.
Example
*IDN?
Response: SANTEC,MPM-210,00000000,Ver.2.0
3. ERR?
Description
Check Error information.
Syntax
ERR?
Response
<value>,<string>
<value>: Code for Error
<string>: Summary of content for Error
Please refer to “5.5.4 ERR?”
Example
ERR?
Response: 0, “No Error”
5-5-2 System-related commands
4. IDIS?
Description
Check recognition of MM for MPM-210.
Syntax
IDIS?
Response
<value0>, <value1>, <value2>, <value3>, <value4>
0: Module is not recognized.
1: Module is recognized.
Example
IDIS?
Response: 1,1,1,1,1