5-8
MPM-210 Operation Manual
6. TRIG
Description
If measuring mode is ‘CONTx’, it checks for measuring
trigger.
Syntax
TRIG<wsp><value1>
Parameters
<value1>
0 – Internal trigger
1 – External trigger
If it is internal trigger, it generates trigger every 10ms
Default
1
Example
TRIG 1
7. TRIG?
Description
If measuring mode is ‘CONTx’, it checks for measuring
trigger.
Syntax
TRIG?
Response
<value1>
0 – Internal trigger
1 – External trigger
Example
TRIG? 1
Response: 1
8. WMOD
Description
Set measurement mode.
Syntax
WMOD<wsp><value1>
Parameters
<value1>
“CONST1” => Constant Wavelength, No Auto Gain,
SME mode
“SWEEP1” => Sweep Wavelength, No Auto Gain,
SME mode
“CONST2” => Constant Wavelength, Auto Gain, SME
mode
“SWEEP2” => Sweep Wavelength, Auto Gain, SME
mode
“FREERUN” => Constant Wavelength, No Auto Gain,
First Hardware Trigger Start (CEM mode)
Refer to the 5.6.1 WMOD.
Default
CONST1
Example
WMOD CONST1
5-5-3 Measurement-related commands