EasyManua.ls Logo

Santec MPM-210 - Page 31

Default Icon
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Multi-Port Optical Power Meter MPM-210
5-9
9. WMOD?
Description
Check measurement mode.
Syntax
WMOD?
Response
“CONST1” => Constant Wavelength, No Auto Gain,
SME mode
“SWEEP1” => Sweep Wavelength, No Auto Gain,
SME mode
“CONST2” => Constant Wavelength, Auto Gain, SME
mode
“SWEEP2” => Sweep Wavelength, Auto Gain, SME
mode
“FREERUN” => Constant Wavelength, No Auto Gain,
First Hardware Trigger Start (CEM mode)
Refer to the 5.6.1 WMOD.
Example
WMOD?
Response: CONST1
10. WAV
Description
Set up for using wavelength of Constant wavelength
Measurement Mode (CONST1, CONST2).
Syntax
WAV<wsp><value1>
Parameters
<value1>
1250.000 ~ 1630.000 Real number
Default
1550
Example
WAV 1550
11. WAV?
Description
Check using wavelength of Constant wavelength
Measurement Mode (CONST1, CONST2).
Syntax
WAV?
Response
<value1>
1250.000 ~ 1630.000 Real number
Unit: nm
Example
WAV?
Response: 1550