5-10
MPM-210 Operation Manual
12. WSET
Description
Set up for using wavelength of Sweep Wavelength
Measurement Mode (SWEEP1, SWEEP2).
Syntax
WSET<wsp><value1>,<value2>,<value3>
Parameters
<value1>: Start wavelength – 1250 ~ 1630
<value2>: Stop wavelength – 1250 ~ 1630
<value3>: Step wavelength – 0.002 ~ 10
Unit: nm
Note) Stop wavelength = Start wavelength + step
wavelength
Default
value1: 1250, value2: 1630, value3: 0.005
Example
WSET 1520,1570,0.05
13. WSET?
Description
Check using wavelength of Sweep Wavelength
Measurement Mode (SWEEP1, SWEEP2).
Syntax
WSET?
Response
<value1>,<value2>,<value3>
<value1>: Start wavelength – 1250 ~ 1630
<value2>: Stop wavelength – 1250 ~ 1630
<value3>: Step wavelength – 0.002 ~ 10
Example
WSET?
Response: 1520,1570,0.05
14 SPE
Description
Set up Wavelength Sweep Speed
Syntax
SPE<wsp><value1>
Parameters
<value1>: Sweep speed – 0.001 ~ 100
Unit: nm/sec
Default
value1: 1
Example
SPE 1