Multi-Port Optical Power Meter MPM-210
5-15
32. LOGN
Description
Set up measurement data point in CONST1/CONST2/
FREERUN measuring mode.
Refer to the 5.6.3 Measurement data setting (LOGN).
Syntax
LOGN<wsp><value1>
Parameters
<value1>: 1 ~ 1,000,000
Default
value1: 1
Example
LOGN 100
34. LOGG?
Description
Read out the logging data.
Syntax
LOGG? <wsp><value1>,<value2>
value1: module number
value2: port number
Response
Please refer to ‘5.6.4 LOGG?’ for detail information.
Example
LOGG?
Response: Please refer to ‘5.6.4 LOGG?’ for detail
information.
33. LOGN?
Description
Check measurement data point.
Refer to the 5.6.3 Measurement data setting (LOGN).
Syntax
LOGN?
Response
<value1>: 1 ~ 1,000,000
Example
LOGN?
Resopnse: 100
31. STAT?
Description
Check measuring status.
Syntax
STAT?
Response
<value1>,<value2>
Value1:
0 – Measuring is still in process.
1 – Measurement completed.
-1 – Setting up for measuring, Force to stop the
measurement.
Value2: Measuring time.
Example
STAT?
Response: 1,100