5-14
MPM-210 Operation Manual
28. CWAVPO?
Description
Power calibration value of the wavelength of “CWAV?”
command Index.
Unit: dB
Refer to the 5.5.5 for the method of wavelength
compensation.
Syntax
CWAVPO?<wsp><module>,<ch>,<index>
Parameters
<module>
Module Number: 0,1,2,3,4 - Integer
<ch>
Port number: 1,2,3,4 - Integer
<index>
Index : Wavelength set order that would be applied –
Integer 1,2,3~ 18,19
Default
<optical power offset> - oat
Example
CWAVPO? 0,1,1
Response: 0.904640
29. MEAS
Description
Command to start measuring.
Syntax
MEAS
Example
MEAS
30. STOP
Description
Command to stop measuring.
Syntax
STOP
Example
STOP
27. CWAV?
Description
Check the wavelength that should be calibrated.
Unit: nm
Refer to the 5.5.5 for the method of wavelength
compensation.
Syntax
CWAV?<wsp><module>,<index>
Parameters
<module>
Module Number: 0,1,2,3,4
<index>
Index : Wavelength set order that would be applied
– Integer 1,2,3~ 18,19,20
Default
<wavelength>
Example
CWAV? 0,1
Response: 1250