Multi-Port Optical Power Meter MPM-210
5-13
19. READ?
Description
Execute optical power or electrical current measurement
and check the result. If sending the “READ” command
during logging mode, logging mode stops.
Syntax
READ?<wsp><module>
Parameters
<module>
Module Number: 0,1,2,3,4
Default
<value1>,<value2>, <value3>,<value4>
value1; Optical power of port 1
value2; Optical power of port 2
value3; Optical power of port 3
value4; Optical power of port 4
Example
READ?
Response: -20.123,-20.454,-20.764,-20.644
23. UNIT?
25. AUTO?
Description
Check measuring unit of optical power or electrical current.
Syntax
UNIT?
Response
<value1>
0 – dBm, dBmA
1 – mW,mA
Example
UNIT?
Response: 0
Description
Check the power range (Auto or Manual).
Syntax
AUTO?
Response
0 – Manual range
1 – Auto range
Example
AUTO?
Response: 1
24. AUTO
Description
Set the power range (Auto or Manual).
It is available for the “READ?” command.
Syntax
AUTO<wsp><value1>
Parameters
<value1>
0 – Manual range
1 – Auto range
Default
value1: 1
Example
AUTO 1