19. AVG?
Description
Check Average time.
Syntax
AVG?
Response
<value1>
value1: 0.01 ~ 10000.00
Unit: ms
resolution: 0.01(10us)
Example
AVG?
Response: 5
20. FGSAVG
Description
Set Average Time.(This command is same operation
as “AVG”.)
Syntax
FGSAVG<wsp><value1>
Parameters
<value1>: 0.01 ~ 10000.00
Unit: ms
Resolution: 0.01(10us)
Default
value1: 0.1
Example
FGSAVG 0.1
21. FGSAVG?
Description
Check Average time. (This command is same
operation as AVG?”.)
Syntax
FGSAVG?
Response
<value1>
value1: 0.01 ~ 10000.00
Unit: ms
resolution: 0.01(10us)
Example
FGSAVG?
Response: 0.1
22. UNIT
Description
Set measuring unit of optical power or electrical current.
Syntax
UNIT<wsp><value1>
Parameters
<value1>
0 – dBm, dBmA
1 – mW, mA
Default
value1: 0
Example
UNIT 0