List Function for Differential Weighing
The list function has 4 display pages: one each for lots, samples, values and
results.
Display Page for Lots
The display page for lots shows all of the lots that have already been created,
as well as the number of samples in each lot and the processing status of the
selected sample (tare, initial and backweighing). On this display page you
can create, rename, delete and print lots. You can also define a factor for
calculation of loss; for instance, to have weight per unit area calculated (such
as grams per square meter). You can also enter a lot ID alphanumerically to
access a lot directly.
Display Page for Samples
This display page shows the samples contained in a selected lot, as well as the
processing status of the samples (tare, initial and backweighing) and the
sample IDs. You can also enter a sample ID alphanumerically to access
a sample directly.
Display Page for Values
This display page shows the date and time of sampling, as well as the sample
ID and the values measured, for a selected sample.
Display Page for Results
This display page shows the calculated values for a selected sample. These
include backweighed residue, loss, loss calculated using a factor, and the ratio
values. The o symbol indicates the value that is selected for display immediately
following a backweighing procedure. To change this setting, use the q and Q
soft keys to move the highlight bar to the desired value, and press l to confirm.
Display Page for Statistics
This page shows the characteristic data for a lot (date; time; statistics on, for
example, the backweighed residue; number of samples) as well as the
calculated values (mean value, standard deviation).
To select a set of statistics from a lot with different numbers of backweighing
procedures:
Press the l soft key to display the selected set of statistics:
oo DeleteCreate QqSample
LOTS: 792 Smpl.avail.
1 1 Sample T
122 1 Sample T,N
AB05 20 Samples T,N,R1
CH0001 10 Samples T,N
CH01234 2 Samples T,N,R1
oo Delete oQqValues
SMPL: avail.792 Lot: CH0001
Sample 1: T,N,R(1) CX87
Sample 2: T,N,R(1) CX88
Sample 3: T,N
Sample 4: T,N
Sample 5: T,N
oo Result oQq
VALUES: Lot: CH0001 Smpl:2
Date,time: 16.11.1998 15:11:17
Name: ID CX88
Tare: T1 + 324.72 g
Net initial wt: N1 + 414.45 g
BackwghId res: R (1)+ 393.55 g
oo Values oQql
RESULT: Lot: CH0001 Smpl:2
Residue: R + 20.74 g
dResidue: R + 80.48 %
Loss: D - 5.03 g
Loss: D - 19.52 %
Ratio1: DR + 24.25 %
STATISTICS: Lot:CH6789
Statistics on: R (1) 5 Spls
Statistics on: R (2) 3 Spls
Statistics on: R (*) 8 Spls
oo Messw. o Qql
STATISTICS: Lot:CH6789
Date,time: 04.02.1999 14:31:30
Statistics on: R (1) OResidueo
No.of values: n 2
Mean value: Mean + 93.28 %
Std. deviation: s 0.01 %
oo Messw. oQql

Table of Contents

Other manuals for Sartorius Factory Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Factory Series and is the answer not in the manual?

Sartorius Factory Series Specifications

General IconGeneral
BrandSartorius
ModelFactory Series
CategoryScales
LanguageEnglish

Related product manuals